Brazilian Journal of Anesthesiology
https://bjan-sba.org/article/doi/10.1590/S0034-70942001000500007
Brazilian Journal of Anesthesiology
Clinical Information

Parestesia no território do nervo mediano: efeito adverso do estimulador de nervo periférico? Relato de caso

Paresthesia at the median nerve territory: adverse effect of the peripheral nerve stimulator? Case report

Carlos Henrique Viana de Castro; Giovanni Menezes Santos; Niwton Carlos Toledo; José Roberto de Rezende Costa

Downloads: 0
Views: 1067

Resumo

JUSTIFICATIVA E OBJETIVOS: A Anestesiologia tem-se desenvolvido continuamente e vem sucessivamente melhorando suas técnicas, fármacos bem como seus equipamentos especialmente aqueles para monitorização. Dentre estes, o estimulador de nervo periférico destaca-se para avaliação adequada do nível de relaxamento neuromuscular e sua recuperação. Objetivou-se mostrar neste relato que, embora a monitorização moderna seja importante, por vezes, pode não ser inócua, gerando suspeição sobre sua suposta ausência de efeitos adversos. RELATO DO CASO: Uma paciente de 56 anos foi submetida a cirurgia estética mamária. Seus antecedentes continham relato de diabetes mellitus, hipotireoidismo, hipertensão arterial, dislipidemia e obesidade, todos com bom controle clínico, medicamentoso e laboratorial, sem menção de complicações anestésicas em cirurgias anteriores. O procedimento transcorreu sem intercorrências e ao seu término, a paciente foi encaminhada à sala de recuperação pós-anestésica. Nesta, antes de receber alta, apresentou queixa importante de parestesia no território do nervo mediano da mão esquerda, ou seja, exatamente no local onde ficou o estimulador de nervo periférico. Tal queixa se reduziu nas próximas 48 horas, tendo a paciente recebido alta hospitalar, sem outros problemas ou alterações do exame físico. CONCLUSÕES: A monitorização com o estimulador de nervo periférico é importante e deve ser preconizada. Para tanto, deve-se fixar e posicionar o membro adequadamente. Não obstante, esta posição pode, em alguns casos, trazer quadros neurológicos compatíveis com compressões nervosas agudas, no pós-operatório. Desta forma, deve-se atentar para diagnósticos diferenciais e, neste caso, o efeito adverso possivelmente não foi ocasionado pelo estimulador de nervo e, sim, pelo posicionamento do membro superior esquerdo.

Palavras-chave

COMPLICAÇÕES, Neurológica, EQUIPAMENTOS

Abstract

BACKGROUND AND OBJECTIVES: Anesthesiology is developing and improving techniques, drugs and equipment, especially monitoring devices. The peripheral nerve stimulator has been used to adequately evaluate muscle relaxation and recovery. This report aimed at showing that, although the importance of modern monitoring, it may sometimes not be innocuous, thus generating suspicion about its supposedly lack of side-effects. CASE REPORT: A fifty-six year-old woman was submitted to cosmetic mammaplasty. Her previous history included: diabetes mellitus, hypothyroidism, hypertension and obesity, all under good clinical control, and with no previous anesthetic complications. Surgery was performed without intercurrences and at the end patient was transferred to PACU. Before being discharged, patient complained of severe paresthesia exactly on the territory of the median nerve where the peripheral nerve stimulator had been placed. This complaint disappeared after 48 h and patient was discharged in good conditions. CONCLUSIONS: Monitoring with peripheral nerve stimulator is important and should be recommended. For such, one must adequately fix and position the limb. However, such position may, in some cases, lead to neurological complications compatible with postoperative nervous compression. So, one must be alert to differential diagnosis and, in our case, it seems that the adverse effect was not caused by the nerve stimulator but by the positioning of the left upper limb.

Keywords

COMPLICATIONS, Neurological, EQUIPMENTS

References

Rudis MI, Sikora CA, Angus E. A prospective, randomized, controlled evaluation of peripheral nerve stimulation versus standard clinical dosing of neuromuscular blocking agents in critically ill patients.. Crit Care Med. 1997;25:575-583.

Zarowitz BJ, Rudis MI, Lai K. Retrospective pharmaco-e conomic evaluation of dosing vecuronium by peripheral nerve stimulation versus standard clinical assessment in critically ill patients.. Pharmacotherapy. 1997;17:327-332.

Baillard C, Gehan G, Reboul-Marty J. Residual curarization the recovery room after vecuronium. Br J Anaesth. 2000;84:394-395.

Jobe MT, Wright PE. Peripheral Nerve Injuries em Cannale: Campbell's Operative Orthopaedics. 1998:3685-3692.