Inappropriate fixation of an endotracheal tube causing cuff malfunction resulting in difficult extubation

Fixação inadequada de tubo endotraqueal, resultando em mau funcionamento do manguito e extubação difícil

Deb Sanjay Nag; Devi Prasad Samaddar

Downloads: 0
Views: 612

Abstract

Abstract We discuss a case of difficult extubation, due to inadequate deflation of the tracheal tube cuff, despite collapse of the pilot balloon, on its aspiration. This was caused by inadvertent kinking of the pilot balloon tubing due to inappropriate tape fixation of the endotracheal tube.

Keywords

Endotracheal, Tube, Fixation, Cuff, Difficult extubation

Resumo

Resumo Relatamos o caso de extubação difícil devido à desinsuflação inadequada do cuff do tubo traqueal, a despeito do colapso do balão piloto, em sua aspiração. Isso foi causado pela torção inadvertida do tubo com balão piloto devido à fixação inadequado do tubo endotraqueal com fita adesiva.

Palavras-chave

Endotraqueal, Tubo, Fixação, Manguito, Extubação difícil

References

Nakagawa H, Komatsu R, Hayashi K. Fiberoptic evaluation of the difficult extubation. Anesthesiology. 1995;82:785-6.

Dryden DE. Circulatory collapse after pneumonectomy (an unusual complication from the use of a catheter): case report. Anesth Analg (Cleve). 1977;56:451-2.

Hulme J, Agarwal S. Failure to deflate an endotracheal cuff. Acta Anaesthesiol Scand. 2008;52:719.

Tanski J, James RH. Difficult extubation due to a kinked pilot tube. Anaesthesia. 1986;41:1060.

Suman S, Ganjoo P, Tandon MS. Difficult extubation due to failure of an endotracheal tube cuff deflation. J Anaesthesiol Clin Pharmacol. 2011;27:141-2.

Hartley M, Vaughan RS. Problems associated with tracheal extubation. Br J Anaesth. 1993;71:561-8.

Guntupalli KK, Bouchek CD. Cricothyroid puncture of an undeflatable endotracheal tube cuff. Crit Care Med. 1984;12:924.

Shamsai J. A new technique for removal of endotracheal tube. Anesth Analg. 2006;103:1040.

Wax D. Where is the evidence?. Anesth Analg. 2007;105:285.

Adhikary SD, George SP, Korula G. Failure of endotracheal cuff deflation. Acta Anaesthesiol Scand. 2005;49:590.

Murdoch E, Holdgate A. A comparison of tape-tying versus a tube-holding device for securing endotracheal tubes in adults. Anaesth Intensive Care. 2007;35:730-5.

5dcd68ca0e88256764bf58f2 rba Articles
Links & Downloads

Braz J Anesthesiol

Share this page
Page Sections